Freestyle – AirSep MN131-1 B User Manual
Page 244

EN
FreeStyle™
EN-32
AirSep
®
Corporation
MN131-1 rev. B - 02/13
Status
Audible
Alarm
Light
Indicates
Action
Alarm
Continuous:
Beep
(Red) alarm;
continuous light
No breath detected
by the unit for a
predetermined time
period.
Check the cannula
connection. Ensure that
you are breathing
through your nose. If the
alarm persists, contact
your Equipment
provider.
Alarm
Intermittent:
Beep, beep,
beep…
25% (Yellow)
BATT; battery
indicator
Battery voltage is
too low to operate
FreeStyle.
Connect the FreeStyle
unit into a DC outlet or
an AC outlet
immediately.
Alarm
Rapid
intermittent:
Beep, beep,
beep…
(Red) alarm;
intermittent light
Breathing rate is
exceeding the
capacity of the
FreeStyle unit.
Reduce activity, then if
necessary use another
source of oxygen as
available. Contact your
Equipment Provider.
Alarm
Rapid
intermittent:
Beep, beep,
beep…
(Red) alarm;
continuous light
General malfunction
of the FreeStyle unit
has occurred.
Turn off the unit.
Change to another
source of oxygen, and
contact your Equipment
Provider.