Calibrating the sensors, Setup calibrating the sensors, Caution – ROTRONIC LOG-HC2 Universal Datalogger User Manual
Page 37

37
Setup
Calibrating the sensors
The sensors and analogue inputs can be calibrated using a 2 point calibration
process . Those sensors marked “not changeable” in the
Calibration column
cannot be calibrated .
CAUTION
Calibration and adjustment of the HygroClip2
probe is carried out using the ROTRONIC
HW4 software .
Select
User settings and at the bottom click on Read user settings . Under
General > Calibration, check the Change values checkbox .
Choose
Channel 3 .
Select a channel and then click
Calibration .
3
4
Calibration can be carried out a) manually or b) by measuring the signal and sub-
sequently entering the target value .
a) Manual calibration
First enter the actual value and then the associated target value for points P1 and
P2 . Activate the completed calibration by checking the checkbox 4 . Confirm
your selection with
OK .
b) Calibration using the measurement function
Apply the desired signal for point P1 and click
Measure . The
measured actual value is displayed . Enter the associated tar-
get value for point P1 . Repeat this procedure for P2 . Activate
the completed calibration by checking the checkbox 4 . Con-
firm your selection with
OK .
Activate the completed calibration by checking the checkbox 4 and then click
OK . If necessary, calibrate further sensors (channels) . With Write user settings
2
you can simultaneously transfer all of the completed calibrations to the logger .