Instructions

18

Operating, transport and storage conditions

• Protect the logger from excessive exposure to the sun and other sources of

heat . Avoid heavy impacts .

• Do not place heavy objects on top of the logger .

• Only store the logger in a dry, dust-free environment .

Operating conditions:

Temperature:

-20°C to +65 °C

Pressure:

500 mbar to 2500 mbar absolute

Humidity:

10-95 % relative humidity, non-condensing

Avoid contact with water and humidity

Optimal storage and transportation conditions:

Temperature:

5 °C bis 45 °C (ideal storage condition for the battery)

Humidity:

10-95 % relative humidity, non-condensing

Avoid contact with water and humidity

Battery charge level:

Never store the logger with a discharged battery .

The ideal charge level is a 2/3 charge

(see „Charging the battery“ on page 15)

Troubleshooting

Problem

Possible cause

Possible solution

The temperature

increases con-

tinuously when

the logger is

connected to the

mains adapter or

the PC

The battery warms up

because charging com-

mences .

When making temperature

measurements disconnect

the logger from mains adapter

or the PC .

The LOG PC

programs do not

recognise the

connected logger

The COM port driver is not

installed correctly .

Observe the manufacturer’s

installation instructions .

http://www .ftdichip .com/