Change basic parameter, Changing the basic unit, Setup change basic parameter – ROTRONIC LOG-HC2 Universal Datalogger User Manual
Page 36

36
Setup
Change basic parameter
Change basic parameter is used to adapt the factory settings (name, unit,
display format) to your requirements . These details are used among others, for
calibration purposes .
Select the desired channel (e .g . analogue input A2) and then click
Change ba-
sic parameter . Enter your adjustments 5 (e .g . Mbar instead of Volt) .
To enter the conversion factor (e .g . Volts to Mbar) the following options are avail-
able:
• Enter the
amplification and the zero point
• Using the calculation assistant 6 enter the values for points P1 and P2 7 .
The program will then calculate the amplification and the zero point .
Activate
Apply user information by checking the checkbox 4 and then click
OK . Transfer the changes to the logger by clicking Write user settings .
Note: Should you wish to use a different display unit to that used for calibration,
then please use the function
Change display parameter .
3
4
6
6
Max . 4 characters
Changes have been made to this channel .
7
5
Resetting channel properties to the factory settings:
Uncheck checkbox 4 (
Apply user information) and transfer the changes to the
logger by clicking
Write user settings .