Limited-Tilt Calibration

Recommended calibration method when >5° of tilt calibration is available, but tilt

is restricted to <45°. (i.e. full range calibration is not possible.)

Accel-Only Calibration

Select this when an accelerometer calibration will be performed.

Accelerometer and Magnetic Calibration (110

d

= 6E

h

)

Selected when magnetic and accelerometer calibration will be done

simultaneously.

Below is a complete sample packet to start a 2D Calibration (kStartCal):

Heading, pitch and roll information is output via the kGetDataResp frame during the

calibration process. This feature provides guidance during the calibration regarding

calibration sample point coverage. During calibration, in the kGetDataResp frame,

the number of data components is set to be 3 and then followed by the data

component ID-value pairs. The sequence of the component IDs are kHeading, kPitch

and kRoll.

kTakeUserCalSample (frame ID 31

d

)

This frame commands the TRAX to take a sample during user calibration. The frame

has no payload.

kUserCalSampleCount (frame ID 17

d

)

This frame is sent from the TRAX after taking a calibration sample point. The

payload contains the sample count with the range of 1 to 18.

kStopCal (frame ID 11

d

)

This command aborts the calibration process. The prior calibration results are

retained.

SampleCount#

UInt32

Payload

00 09

Frame ID

0A

ByteCount

00 00

CalOption

(MSBs)

CalOption

(2D Calibration)

00 14

5C F9

Checksum