Table 5-2: 12 Point Full-Range Calibration Pattern

Sample #

Yaw

1

Pitch

Roll

First Circle

1

±5°

30° to 40°

2

90°

±5°

-30° to -40°

3

180°

±5°

30° to 40°

4

270°

±5°

-30° to -40°

Second Circle

5

30°

> +45°

30° to 40°

6

120°

> +45°

-30° to -40°

7

210°

> +45°

30° to 40°

8

300°

> +45°

-30° to -40°

Third Circle

9

60°

< -45°

30° to 40°

10

150°

< -45°

-30° to -40°

11

240°

< -45°

30° to 40°

12

330°

< -45°

-30° to -40°

Footnote:

1. Yaw listings are not absolute heading directions but rather relative heading

referenced to the first sample.

5.1.2

2D Calibration

A 2D Calibration is intended for very low tilt operation (<5

°) where calibrating the

TRAX with greater tilt is not practical.

This procedure calibrates for hard and soft-iron effects in only two dimensions, and in

general is effective for operation and calibration in the tilt range of -5

° to +5°. The

recommended 12 point calibration pattern is a circle of evenly spaced points, as given in

Table 5-3.