beautypg.com

3 calibration results, Mag cal, Accel cal – PNI FieldForce Trax User Manual

Page 33: Distribution error, Calibration results

background image

PNI Sensor Corporation

DOC#1016505 r06

TRAX User Manual

Page 29

If “Automatic Sampling” is checked the TRAX needs to be held steady for a short

time and then a sample automatically will be taken. Once the window indicates

the next number, the device’s orientation should be changed and held steady for

the next sample. Once the pre-set number of samples has been taken (as set on

the Configuration tab) the calibration is complete.

Regardless of whether “Automatic Sampling” is selected, two criteria must be met for a
calibration sample to be taken. First, the TRAX must be held steady enough to meet

PNI-defined stability criteria. Second, the TRAX’s orientation must have changed

enough to meet PNI-defined orientation change criteria.

If “H/P/R Output During Cal” is selected, then the “Heading”, “Pitch”, and “Roll” fields
will be populated in real-time once the button is selected. Note that the readings

in these fields are relative, since calibration is in process.

6.5.3

Calibration Results

Once the calibration is complete the “Calibration Results” section will indicate the

quality of the calibration. This may take a few seconds to populate. The primary score

of concern is the MagCalScore or AccelCalScore, depending on whether a magnetic or

accelerometer calibration was performed. The other parameters provide information that

may assist in improving the MagCalScore should it be unacceptably high. If a calibration

is acceptable, then click the button to save the calibration coefficients to the

coefficient set defined on the Configuration Tab.

Note: If a calibration is aborted, all the score’s will read “179.80”, and the calibration coefficients
will not be changed. (Clicking the button will not change the calibration coefficients.)

Mag Cal

Represents the over-riding indicator of the quality of a magnetic calibration.

Acceptable scores are <1 for Full-Range Calibration and <2 for other methods. Note

that it is possible to obtain acceptable Distribution Error and Tilt Error scores and still

have a rather high Mag Cal value. The most likely reason for this is the TRAX is

close to a source of magnetic distortion that is not fixed with respect to the device.

Accel Cal

Represents the quality of an Accelerometer Calibration. The score should be <1.

Distribution Error

Indicates if the distribution of sample points is good, with an emphasis on the heading

distribution. The score should be 0. Significant clumping or a lack of sample points