Note: While the TRAX is shown removed from the host system, the Accelerometer
Calibration may be performed with the TRAX mounted in the host system.

Figure 5-2: Accelerometer Calibration Starting Orientations

5.2.2

Mag-and-Accel Calibration

The TRAX allows for a simultaneous magnetometer and accelerometer calibration. This

requires a good calibration pattern, stable measurements (not handheld), and installation
in the user’s system such that the appropriate local magnetic environment is present. PNI
recommends 18 to 32 calibration points for a Mag-and-Accel Calibration. The Accel-

Only Calibration pattern discussed in Section 5.2.1 also will work for a Mag-and-Accel

Calibration. Optimal performance is obtained when all rotations of the cube are

performed towards magnetic north to achieve the widest possible magnetic field

distribution.

Note that combining calibrations only makes sense if all the host system’s magnetic
distortions (steel structures or batteries, for instance) are present and fixed relative to the

module when calibrating. If the Accel-Only Calibration is performed, the user’s system

distortions are not relevant, which allows the TRAX to be removed from the host system

in order to perform the Accel-Only Calibration.