beautypg.com

Declination, Filter taps, Flush filter – PNI FieldForce Trax User Manual

Page 29: 2 acquisition settings, Acquisition mode, Acquisition settings

background image

PNI Sensor Corporation

DOC#1016505 r06

TRAX User Manual

Page 25

Declination

The declination represents the heading difference between magnetic north and true

north, and needs to be entered if “True” is selected as the “North Reference”.

Declination varies primarily with location, although it also gradually changes over

time (years) for a given location. To find the declination for where the TRAX will be

used, go to

http://www.ngdc.noaa.gov/geomagmodels/Declination.jsp

.

Filter Taps

When operating the TRAX in Compass Mode, the TRAX will use a finite impulse

response (FIR) filter to effectively provide a stable heading reading. When operating

in AHRS Mode a completely different filtering method is used and the number of FIR

Taps is not relevant.

The number of taps (or samples) represents the amount of filtering to be performed.

Either 0, 4, 8, 16, or 32 taps may be selected, with zero taps representing no filtering.

Note that selecting a larger number of taps can significantly slow the time for the
initial sample reading and, if “Flush Filters” is selected, the rate at which data is
output. The default setting is 32.

The FIR filtering operates on a first-in, first-out (FIFO) basis. For example, with FIR

Taps set to 32, once the initial 32 samples are taken the next sample (#33) is included

in the filtering and the initial sample (#1) is dropped from the filtering.

Flush Filter

The Flush Filter setting is only relevant when in Compass Mode, as it affects how the

FIR filters are implemented. Flushing the FIR filter clears all the filter values so it is

necessary to fully repopulate the filter before a good reading can be given. For

example, if 32 FIR taps is set, then 32 new samples must be taken to provide a good

reading. It is particularly prudent to flush the filter if the Sampling Delay is set to a

non-zero value as this will purge old data. Note that flushing the filters increases the

delay until data is output, with the length of the delay being directly correlated to the

number of FIR taps. The default is not to Flush Filters.

6.4.2

Acquisition Settings

Acquisition Mode

When operating in Continuous Acquisition Mode, the TRAX continuously outputs

data to the host system. The rate is set by the Sample Delay. When operating in

Polled Mode, TRAX Studio simulates a host system and polls the TRAX for a single

measurement; but TRAX Studio makes this request at a fixed rate which is set by the

Polling Delay. In both cases data is continuously output, but in Continuous