beautypg.com

5 user calibration, User calibration, 5user calibration – PNI FieldForce Trax User Manual

Page 16

background image

PNI Sensor Corporation

DOC#1016505 r06

TRAX User Manual

Page 12

5

User Calibration

The magnetic sensor in the TRAX is calibrated at PNI’s factory in a magnetically controlled

environment. However sources of magnetic distortion positioned near the TRAX in the user’s
system will distort Earth’s magnetic field and should be compensated for in the host system with
a user calibration. Examples of such sources include ferrous metals and alloys (ex. iron, nickel,

steel, etc.), batteries, audio speakers, current-carrying wires, and electric motors. Compensation

is accomplished by mounting the TRAX in the host system and performing a user calibration. It
is expected the sources of magnetic distortion remain fixed relative to the TRAX‘s position
within the host system. By performing a calibration, the TRAX identifies the local sources of

magnetic distortion and negates their effects from the overall reading to provide an accurate

heading.

As with the magnetic sensor, the accelerometer in the TRAX is calibrated at PNI’s factory. But

the accelerometer gradually changes over time, and the user either will need to periodically

perform a user accelerometer calibration or return the unit to PNI for recalibration. As a rule-of-

thumb, the accelerometer should be recalibrated every 6 to 12 months. Unlike a magnetic

calibration, the accelerometer may be calibrated outside the host system. Accelerometer

calibration is more sensitive to noise or hand jitter than magnetic calibration, especially for

subsequent use at high tilt angles. Because of this, a stabilized fixture is suggested for

accelerometer calibration, although resting the unit against a stable surface often is sufficient.

Key Points:

Magnetic calibration:

o Requires incorporating the TRAX into the user’s host system such that the

magnetic components of the user’s system can be compensated for.

o Allows for 4 different methods of calibration. Full-Range Calibration provides

the highest heading accuracy, but requires

≥45° of pitch. 2D and Limited-Tilt

Calibration allow for good calibration when the range of allowable motion is

limited. Hard-Iron-Only Calibration updates the hard-iron compensation

coefficients with a relatively easy procedure.

Accelerometer calibration requires rotating the TRAX through a full sphere of coverage,

but it does not need to be incorporated in the user’s system during calibration.

If the TRAX will experience different states during operation, such as operating with a

nearby shutter sometimes closed and sometimes open, or operating over a broad

temperature range, then different sets of calibration coefficients can be saved for the

various states. Up to 8 magnetic calibration coefficient sets and 3 accelerometer

calibration coefficient sets can be saved.