CONFIGURATION MENU
DO NOT ENTER THE CONFIGURATION MENU UNLESS YOU HAVE
BEEN INSTRUCTED TO BY THESE INSTRUCTIONS. INCORRECT
ENTRY OF DATA IN THE CONFIGURATION MENU MAY CAUSE
IMPROPER OR UNPREDICTABLE OPERATION OF THE INSTRUMENT.
The Configuration Menu is used to quickly recover the instrument after certain Error
Codes. The configuration for your particular model is shown on the Model / Serial label
located on the left side (when viewed from the front) of the instrument housing. A label
found inside on the right side of the chassis shows the same information.
If you do recover the instrument from a F A i L t E S t error, the memory configuration is
restored to the factory settings. All hardware inputs and outputs are automatically
recognized except for Options 924, 926, and 928.
If you have a instrument equipped with one of these options, Factory Default will not
automatically restore these settings. They must be set by hand.
To re-configure:
1.
At power up, simultaneously press and hold the INDEX & ENTER keys while the lamp
test or self test is displayed. Hold the keys down until the ROM ID code appears.
2.
Press the INDEX key to advance to the next menu item, F A C t d F L t . This function
restores the control to original factory settings. It will also remove all of the hardware,
option, and software configuration values. After selecting F A C t d F L t you will need to
restore the configuration values as shown above to allow the instrument to operate
correctly.
To restore factory default values, press and hold the ENTER key. While holding the
ENTER key also press the DOWN ARROW key. The display will blink momentarily
and the instrument will reboot. The instrument will then display the ROM ID. Press
INDEX to step through F A C t d F L t and continue at step 3.
3. Press INDEX to O P t 1 . If your instrument is equipped with Option 924, 926, or 928,
press the UP ARROW or DOWN ARROW as necessary to display r S P t . The display
will be flashing. To select, press ENTER while r S P t is displayed. When selected, the
option number will stop flashing.
DO NOT SELECT AN OPTION IF THE INSTRUMENT DOES NOT HAVE THE
HARDWARE TO SUPPORT IT. SELECTION OF AN UNSUPPORTED
OPTION MAY CAUSE IMPROPER OR UNPREDICTABLE OPERATION OF
THE INSTRUMENT.