Managing the Database
37
FastCalXP Operation Manual
Use Custom Certifi cate Template
When
Use custom certifi cate
template is checked, you can override the default Excel template defi ned in the Options dialog. This allows diff erent manufactur-
ers or models to have unique certifi cate formats.
Cal Interval
Enter the
Cal interval
(in days) which will be used for calibration scheduling.
Transmitter
Select
Transmitter
when you will select which parameters of the transmitter to measure during calibration.
Measure mA
Select
Measure mA
when calibrating the analog output of the transmitter. Specify the Low/High pressure range as well as the corresponding Low/High mA
output of the transmitter.
Measure HART
If the model is a smart transmitter, select
Measure HART
to view digital output of the transmitter. If the transmitter has an external display check
Measure
Displayed
to include the external display in the certifi cate. You can select either
Measure Displayed
or
Measure HART
but not both for a model.
Accuracy
There are three ways to defi ne the accuracy of the model. Many gauges will have an ASME or ISO/IEC grade which can be selected from the pull-down list. If
the accuracy is not listed, you may be able to simply enter a
% of span
accuracy.
Alternatively, select
Custom
to defi ne a custom accuracy by specifying a series of pressure ranges and accuracies associated with those ranges. Click the
Add
button
to include a range and accuracy in the specifi cation. When the accuracy is fully specifi ed, click
OK
.