Calibrating Gauges

11

FastCalXP Operation Manual

8

Select the

Accuracy

of the gauge from the pull-down list or specify a

% of span

or

Custom

accuracy.

Refer to the discussion of

Models on page 36

in the

Managing the Database section.

9

Click

Next

to select the As Received and As Left calibration Profi les.

If the desired profi les are not listed, click on the

Create New Profi le button

to open the Profi les editor. Refer to the discussion of

Test Profi les on page 39

.

10

Select a location from the pull-down list or click on the

Add New Location button

.

A location is not required. However, it is commonly used to specify the operating location of the gauge within the user’s facility or, by a calibration labora-

tory, for customer information. Use the boxes provided to enter specifi c location information and notes