Calibrating Gauges
15
FastCalXP Operation Manual
Autodetection
FastCalXP automatically detects Crystal Engineering’s nVision Reference Recorders, XP2i (XP2) digital pressure gauges and 30 Series calibrators. Once the
gauge and the reference have been mounted on the GaugeCalHP and the serial or USB cable has been connected to the reference and the computer’s ap-
propriate port, the autodetection and calibration can proceed. Calibration of a smart transmitter with a HART modem requires that the HART modem be
connected to the computer and the HART modem leads be connected to the power leads of the transmitter. Refer to the
Hardware Setup and Connections
section
and t
for help with confi guring your calibration work station.
To Automatically Detect a New or Previously Used Reference
1
Turn the reference on and click the
Perform New As Received Cal button
or the
Perform As Left Cal button
.
FastCalXP will take a few seconds to detect the reference. When a new reference has been detected it can be added to the database and the current calibra-
tion certifi cate information must be entered by the operator.
2
Enter the
Latest calibration certifi cate number
and the date on which that
Certifi cate expires.
If the auto-detected reference exists in the database, FastCalXP will proceed to the Exercise Window or to the Calibration Window.
If the reference is not automatically detected refer to the
.
Adding References Manually
References can be added manually using the Database Manager. Refer to the