Crystal FastCalXP Calibration Software User Manual
Page 20

Calibrating Gauges
18
FastCalXP Operation Manual
To Change the Number of Exercise Cycles Associated with a Profi le
1
choose
Manage Database
from the Tools menu and click on the
Profi les button
.
2
Select the profi le of interest and click on the
Edit button
.
3
Enter the desired number of exercise cycles in the
Number of exercise cycles
text box.
Matching the Target
Two diff erent approaches can be taken while stepping through a calibration profi le: matching the gauge to the Target Pressure or matching the reference to
the Target Pressure.
During a calibration, the
Target Pressure
is the next profi le value and the
Reference
and
Gauge
pressures are indicated below. The Gauge value is, by default, the
same value as the Target Pressure. However, the value can be changed by the user to match what is read from the gauge.
To Match the Gauge to the Target Pressure
1
Read the gauge while turning the handle of the GaugeCalHP.
2
Stop when the gauge reaches the Target Pressure value.
Matching the gauge allows for much faster calibration since interpolation is not required to read the gauge and the Gauge value does not have to be
entered.