Settings and Options
28
FastCalXP Operation Manual
Save Certifi cates To
The calibration certifi cates created by FastCalXP will be saved in the directory specifi ed in
Save certifi cates to.
This directory is set the fi rst time the software is
run after installation or when a new database is created and can be changed here using the Browse button.
Certifi cate Naming Template
The
Certifi cate naming template
is used by FastCalXP to name the calibration certifi cates. In general, certifi cates are numbered with the
Next certifi cate
number
incrementing automatically each time a certifi cate is created. Use number signs (#’s) in the Certifi cate naming template to defi ne the number of digits
to be used.
Calibration Representative and Quality Representative
The
Calibration representative
and the
Quality representative
must be listed users and can be selected using the pull-down lists. In the certifi cate template
provided with the software these names will be listed, with signature lines, on the fi rst page of each certifi cate.
Calibration Settings
The Calibration Settings defi ne the default Calibration Method.
Actual
Uses the pressure reading from the reference at the moment the Next Point button is clicked.
Peak
Automatically advances to the Next Point when a peak is detected.
Head Correction
If the gauge and the reference do not sit at the same height, a
Head correction
can be applied. Enter a correction to be added to the reference pressure, or use
the Correction calculator to calculate a correction given the specifi c conditions of the calibration. Refer to the discussion of the
.