Operation – KROHNE MFC 400 FOUNDATION FIELDBUS User Manual
Page 47
OPERATION
4
47
MFC 400
www.krohne.com
11/2013 - 4003221901 - AD MFC 400 FF R01 en
PRESS_SUPP_
PRESS_SUPP_
PRESS_SUPP_
PRESS_SUPP_
TIME
TIME
TIME
TIME
Press. Supp. Time
R/W
Sets the time that pressure suppression is active (Fct. C1.3.2). 0.0 s
Settings:
0.0…20.0 s
PROD_DENSITY_
PROD_DENSITY_
PROD_DENSITY_
PROD_DENSITY_
VAL
VAL
VAL
VAL
Product Density
Value
R/W
Density of the product for the calibration point (Fct. C1.2.1).
998.2 kg/m³
Settings:
-1000.00…5000.0 kg/m³
REF_DENSITY_
REF_DENSITY_
REF_DENSITY_
REF_DENSITY_
SLOPE
SLOPE
SLOPE
SLOPE
Ref. Density Slope
R/W
Set density slope for referred density (Fct. C1.2.4).
0.0
Settings:
0.0…65.0
SECONDARY_
SECONDARY_
SECONDARY_
SECONDARY_
VALUE
VALUE
VALUE
VALUE
Density
The measured value and status available to the Function
Block.
-
STATUS
Status
R
Digital transducers, unlike their analogue versions, can detect
faults that make the measurement bad or prevent the
actuator from responding. This additional, valuable
information will be passed along with each transmission of a
data value in the form of a status attribute.
-
VALUE
Value
R
A numerical quantity entered by a user or calculated by the
algorithm.
-
SECONDARY_
SECONDARY_
SECONDARY_
SECONDARY_
VALUE_RANGE
VALUE_RANGE
VALUE_RANGE
VALUE_RANGE
Density Range
Defines the ends of the SECONDARY_VALUE, the units of the
SECONDARY_VALUE, and the decimal point position (number
of significant digits to the right of the point). The engineering
units must match the units selected in the parameter
XD_SCALE of the Analog Input Block that reads the channel
with this value.
-
EU_100
max value
R
The engineering unit value which represents the upper end of
range of the associated Block Parameter.
5000 kg/m³
EU_0
min value
R
The engineering unit value which represents the lower end of
range of the associated Block Parameter.
0.05 kg/m³
UNITS_INDEX
Units Index
R/W
Device Description units code index for the engineering unit
descriptor for the associated block value.
kg/m³
Settings:
kg/m³ / kg/dm³ / kg/L / g/m³ / g/L / g/cm³ / g/ml / Mg/m³ /
t/m³ / lb/in³ / lb/ft³ / lb/gal / lb/ImpGal
DECIMAL
Decimal (Point)
R
The number of digits to the right of the decimal point which
should be used by an interface device in displaying the
specified parameter.
0
SECONDARY_
SECONDARY_
SECONDARY_
SECONDARY_
VALUE_TYPE
VALUE_TYPE
VALUE_TYPE
VALUE_TYPE
Secondary Value
Type
R
The type of measurement represented by the secondary value.
For Example: Mass Flow, Absolute Pressure, Differential
Temperature.
Density
Settings:
Density
SENSOR_CAL_
SENSOR_CAL_
SENSOR_CAL_
SENSOR_CAL_
DATE
DATE
DATE
DATE
Sensor Calibration
Date
R
The date of the last sensor calibration.
-
SENSOR_CAL_
SENSOR_CAL_
SENSOR_CAL_
SENSOR_CAL_
LOC
LOC
LOC
LOC
Sensor Calibration
Location
R/W
The last physical location at which the sensor was calibrated.
(ex. Acme Labs)
KROHNE
SENSOR_CAL_
SENSOR_CAL_
SENSOR_CAL_
SENSOR_CAL_
WHO
WHO
WHO
WHO
Sensor Calibration
Who
R/W
The name of the person responsible for the last sensor
calibration.
KROHNE
Parameter
Parameter
Parameter
Parameter
DD name
Access
Description and settings
Initial Value