OPERATION

4

47

MFC 400

www.krohne.com

11/2013 - 4003221901 - AD MFC 400 FF R01 en

PRESS_SUPP_

PRESS_SUPP_

PRESS_SUPP_

PRESS_SUPP_

TIME

TIME

TIME

TIME

Press. Supp. Time

R/W

Sets the time that pressure suppression is active (Fct. C1.3.2). 0.0 s

Settings:

0.0…20.0 s

PROD_DENSITY_

PROD_DENSITY_

PROD_DENSITY_

PROD_DENSITY_

VAL

VAL

VAL

VAL

Product Density

Value

R/W

Density of the product for the calibration point (Fct. C1.2.1).

998.2 kg/m³

Settings:

-1000.00…5000.0 kg/m³

REF_DENSITY_

REF_DENSITY_

REF_DENSITY_

REF_DENSITY_

SLOPE

SLOPE

SLOPE

SLOPE

Ref. Density Slope

R/W

Set density slope for referred density (Fct. C1.2.4).

0.0

Settings:

0.0…65.0

SECONDARY_

SECONDARY_

SECONDARY_

SECONDARY_

VALUE

VALUE

VALUE

VALUE

Density

The measured value and status available to the Function

Block.

-

STATUS

Status

R

Digital transducers, unlike their analogue versions, can detect

faults that make the measurement bad or prevent the

actuator from responding. This additional, valuable

information will be passed along with each transmission of a

data value in the form of a status attribute.

-

VALUE

Value

R

A numerical quantity entered by a user or calculated by the

algorithm.

-

SECONDARY_

SECONDARY_

SECONDARY_

SECONDARY_

VALUE_RANGE

VALUE_RANGE

VALUE_RANGE

VALUE_RANGE

Density Range

Defines the ends of the SECONDARY_VALUE, the units of the

SECONDARY_VALUE, and the decimal point position (number

of significant digits to the right of the point). The engineering

units must match the units selected in the parameter

XD_SCALE of the Analog Input Block that reads the channel

with this value.

-

EU_100

max value

R

The engineering unit value which represents the upper end of

range of the associated Block Parameter.

5000 kg/m³

EU_0

min value

R

The engineering unit value which represents the lower end of

range of the associated Block Parameter.

0.05 kg/m³

UNITS_INDEX

Units Index

R/W

Device Description units code index for the engineering unit

descriptor for the associated block value.

kg/m³

Settings:

kg/m³ / kg/dm³ / kg/L / g/m³ / g/L / g/cm³ / g/ml / Mg/m³ /

t/m³ / lb/in³ / lb/ft³ / lb/gal / lb/ImpGal

DECIMAL

Decimal (Point)

R

The number of digits to the right of the decimal point which

should be used by an interface device in displaying the

specified parameter.

0

SECONDARY_

SECONDARY_

SECONDARY_

SECONDARY_

VALUE_TYPE

VALUE_TYPE

VALUE_TYPE

VALUE_TYPE

Secondary Value

Type

R

The type of measurement represented by the secondary value.

For Example: Mass Flow, Absolute Pressure, Differential

Temperature.

Density

Settings:

Density

SENSOR_CAL_

SENSOR_CAL_

SENSOR_CAL_

SENSOR_CAL_

DATE

DATE

DATE

DATE

Sensor Calibration

Date

R

The date of the last sensor calibration.

-

SENSOR_CAL_

SENSOR_CAL_

SENSOR_CAL_

SENSOR_CAL_

LOC

LOC

LOC

LOC

Sensor Calibration

Location

R/W

The last physical location at which the sensor was calibrated.

(ex. Acme Labs)

KROHNE

SENSOR_CAL_

SENSOR_CAL_

SENSOR_CAL_

SENSOR_CAL_

WHO

WHO

WHO

WHO

Sensor Calibration

Who

R/W

The name of the person responsible for the last sensor

calibration.

KROHNE

Parameter

Parameter

Parameter

Parameter
DD name

Access

Description and settings

Initial Value