(1)(12)(31):Check process for entrained gas. / (1)(13)(31):

Check process cond. In case of remote sensor check cabling. /

(1)(14)(28):Check cable and connection at terminal B, reduce

load. / (1)(14)(29):Check cable and connection at terminal B,

reduce load. / (1)(14)(30):Check cable and connection at

terminal A, reduce load. / (1)(14)(31):Check cable and

connection at terminal A, reduce load. / (1)(15)(30):Check

device connected to current input B. / (1)(15)(31):Check device

connected to current input A. / (1)(16)(00):reserved /

(1)(17)(30):Save settings to Backup 2. / (1)(17)(31):Save

settings to Backup 1. / (1)(18)(29):Change one param. The

status should disappear within 1 minute. / (1)(18)(30):Please

contact the factory. / (1)(18)(31):Check if converter is correctly

mounted and fixed into housing. / (1)(19)(00):reserved /

(1)(20(27):Check process conditions or pipe diameter setting. /

(1)(20)28):Check process conditions and density calibration. /

(1)(20)(30):Check process conditions. / (1)(20)(31):Check

process conditions. / (1)(21)(03):Check the range settings of

the related output. / (1)(21)(04):Check the range settings of the

related output. / (1)(21)(05):Check the range settings of the

related output. / (1)(21)(06):Check the range settings of the

related output. / (1)(21)(07):Check the range settings of the

related output. / (1)(21)(09):Check the range settings of the

related output. / (1)(22)(26):Check reason for electronics over

temperature. / (1)(23)(31):Check cabling in case of remote

device or replace sensor. / (1)(24)(00):reserved /

(1)(25)(07):Turn off simulation / (1)(25)(08):Turn off

simulation / (1)(25)(09):Turn off simulation / (1)(25)(10):Turn

off simulation / (1)(25)(11):Turn off simulation /

(1)(25)(12):Turn off simulation / (1)(25)(22): Check Fieldbus

Configuration. / (1)(25)(28): No Action Required. / (1)(25)(29):

Turn off simulation mode / (1)(25)(30):For normal operation

turn off simulation mode. / (1)(25)(31):For normal operation

turn off simulation mode. / (1)(26)(00):reserved /

(1)(27)(00):reserved / (1)(28)(00):reserved / (1)(29)(02):Check

IO B configuration or reset to factory configuration. /

(1)(29)(03):Check IO B configuration or reset to factory

configuration. / (1)(29)(04):Check IO A configuration or reset to

factory configuration. / (1)(29)(05):Check IO A configuration or

reset to factory configuration. / (1)(29)(10):Check IO2

configuration or reset to factory configuration. /

(1)(29)(11):Check IO1 configuration or reset to factory

configuration. / (1)(29)(12):Check display configuration or

reset to factory configuration. / (1)(29)(21): Check fieldbus

configuration or reset to factory configuration. / (1)(29)(25):

Check config. of the related function. Reset config. to factory. /

(1)(29)(26):Check config. of the related function. Reset config.

to factory. / (1)(29)(30):Check config. of the related function.

Reset config. to factory. / (1)(29)(31):Check config. of the

related function. Reset config. to factory. / (1)(30)(01):Perform

power reset. If the status returns, contact the factory. /

(1)(30)(02):Perform power reset. If the status returns, contact

the factory. / (1)(30)(07):Perform power reset. If the status

returns, contact the factory. / (1)(30)(08):Perform power reset.

If the status returns, contact the factory. / (1)(30)(16):Perform

power reset. If the status returns, contact the factory. /

(1)(30)(21):Perform power reset. If the status returns, contact

the factory. / (1)(30)(26):Perform power reset. If the status

returns, contact the factory. / (1)(30)(27):Perform power reset.

If the status returns, contact the factory. / (1)(30)(28):Perform

power reset. If the status returns, contact the factory. /

(1)(30)(31):Perform power reset. If the status returns, contact

the factory. / (1)(31)(30): Check cabling in case of remote

device or replace sensor. / (1)(31)(31):Replace sensor

FD_SIMULATE

FD_SIMULATE

FD_SIMULATE

FD_SIMULATE

Field Diagnostic

Simulate

Used as the field diagnostic condition when the simulation is

enabled.

-

Parameter

Parameter

Parameter

Parameter
DD name

Access

Description and settings

Initial Value