Operation – KROHNE OPTIWAVE 7300C EN User Manual
Page 62
6
OPERATION
62
OPTIWAVE 7300 C
www.krohne.com
03/2014 - 4000172405 - HB OPTIWAVE 7300 R07 en
B. Test
Menu
No.
Function
Function description
Selection list
Default
B.1 Test
B.1.1
Show Output 1
This displays analogue output 1 value
[mA].
Read only
B.1.2
Set Output 1
This sets analogue output 1 to a test
value [mA] selected from a list. Output
will change to the selected value,
independent of the measured value.
3.6, 4, 6, 8, 10, 12, 14, 16, 18, 20 or
22 mA
4 mA
B.1.3
Show Output 2
This displays analogue output 2 value
[mA].
Read only
B.1.4
Set Output 2
This sets analogue output 2 to a test
value [mA] selected from a list. Output
will change to the selected value,
independent of the measured value.
3.6, 4, 6, 8, 10, 12, 14, 16, 18, 20 or
22 mA
4 mA
B.1.5
Internal Test
This initiates the hardware test. The
device displays the results.
Read only
B.2 Information
B.2.1
Outputs
Analogue output settings. This includes
assigned functions, 4 … 20 mA scale
settings, error handling and HART
®
parameters.
Read only
B.2.2
15 Min Log
A log of output values for the last 15
minutes. A log is taken every 10
seconds and displayed on a graph.
Read only
B.2.3
Device ID
This displays device order no, V-no,
service no, Ex approval, Main CPU
version, Companion CPU version and
DSP version.
Read only
B.2.4
Quick Setup
Sum.
A summary of the parameters entered
in the quick setup menu
Read only
B.2.5
TAG number
The TAG number can be seen and
updated here
?
TAGNO0123
4567890
Temperature
Temperature of the electronics block.
The display will automatically switch
off if the temperature is below -20°C /
-4°F or above +60°C / +140°F.
Read only
B.2.6
Error Records
A log of device errors. Scroll down the
list and push [^
^
^
^] to display the error
details. Opening a log will remove the
error icon if it appeared in operator
mode.
Read only
Measurement
Quality
Status of device errors at this time. A
"tick" symbol next to an error in the list
shows that this error is active and may
have an unwanted effect on the device.
Read only