Description of hart interface, 9 hart, Menu tree for ams – KROHNE OPTIWAVE 7300C EN User Manual
Page 119: Configure/setup
DESCRIPTION OF HART INTERFACE
9
119
OPTIWAVE 7300 C
www.krohne.com
03/2014 - 4000172405 - HB OPTIWAVE 7300 R07 en
9.9 HART
®
menu tree for AMS
Abbreviations of the following tables:
•
Opt
Optional, depending on device version and configuration
•
Rd
Read only
9.9.1 Overview AMS menu tree (positions in menu tree)
9.9.2 AMS menu tree (details for settings)
7 Service Variables
1 Service Parameters
2 Calibration Parameters
3 Maintenance Info
1 Service Number
Rd
/ 2 Service Date
Rd
/ 3 Operator
Rd
1 If it is necessary to change the settings, enter the password. The default password is 123412.
Configure/Setup
Installation Parameters 1
Installation Parameters 2
Output Selection
Primary Output Setting
Secondary Output Setting
Service
HART
Device
Device Diagnostics
Overview
Critical
Informational
Process Variables
Calibration Management
Methods
Access Right
Diagnostics and Test
Calibrate
Tables
Reset
Configure/Setup
Installation Parameters 1
Tank Type / Application Type / Tank Height / Measuring Range
Opt
/
Stillwell Height
Opt
/ Stillwell Diameter
Opt
/ Blocking Distance /
Antenna Extension / Distance Piece / Tank Bottom Offset / Measuring
Mode / Product Er / Tracing Velocity / Multiple Reflections / Empty
Spectrum... / Time Constant / Reference Offset / Overfill Detection
Installation Parameters 2
Length Unit / Volume Unit / Mass Unit / Table Length Unit / Conversion
Unit
Opt
/ Custom. Length Unit / Custom. Length R... / Custom. Conv.
Unit / Product Density
Opt
/ Output2 Status
Opt
/ Conversion Flag
Opt