beautypg.com

Description of hart interface, 9 hart, Menu tree for ams – KROHNE OPTIWAVE 7300C EN User Manual

Page 119: Configure/setup

background image

DESCRIPTION OF HART INTERFACE

9

119

OPTIWAVE 7300 C

www.krohne.com

03/2014 - 4000172405 - HB OPTIWAVE 7300 R07 en

9.9 HART

®

menu tree for AMS

Abbreviations of the following tables:

Opt

Optional, depending on device version and configuration

Rd

Read only

9.9.1 Overview AMS menu tree (positions in menu tree)

9.9.2 AMS menu tree (details for settings)

7 Service Variables

1 Service Parameters

2 Calibration Parameters

3 Maintenance Info

1 Service Number

Rd

/ 2 Service Date

Rd

/ 3 Operator

Rd

1 If it is necessary to change the settings, enter the password. The default password is 123412.

Configure/Setup

Installation Parameters 1

Installation Parameters 2

Output Selection

Primary Output Setting

Secondary Output Setting

Service

HART

Device

Device Diagnostics

Overview

Critical

Informational

Process Variables

Calibration Management

Methods

Access Right

Diagnostics and Test

Calibrate

Tables

Reset

Configure/Setup

Installation Parameters 1

Tank Type / Application Type / Tank Height / Measuring Range

Opt

/

Stillwell Height

Opt

/ Stillwell Diameter

Opt

/ Blocking Distance /

Antenna Extension / Distance Piece / Tank Bottom Offset / Measuring
Mode / Product Er / Tracing Velocity / Multiple Reflections / Empty
Spectrum... / Time Constant / Reference Offset / Overfill Detection

Installation Parameters 2

Length Unit / Volume Unit / Mass Unit / Table Length Unit / Conversion
Unit

Opt

/ Custom. Length Unit / Custom. Length R... / Custom. Conv.

Unit / Product Density

Opt

/ Output2 Status

Opt

/ Conversion Flag

Opt