Stilling wells - general notes
Installation in tanks containing one liquid and foam

• Drill an air circulation hole (max. Ø10 mm / 0.4¨) in the stilling well above the maximum level.
• Remove the burr from the hole.

Installation in tanks containing one liquid or more without foam

• Drill an air circulation hole (max. Ø10 mm / 0.4¨) in the stilling well above the maximum level.
• Drill 1 or more liquid circulation holes in the stilling well (if there is more than 1 liquid in the

tank).

i

These holes help the liquid to move freely between the stilling well and the tank.

• Remove the burr from the hole.

Stilling wells: floating roofs

If the device must be installed on a tank with a floating roof, install it in a stilling well.

Figure 3-10: Floating roofs

1 Sediment
2 Support fixtures
3 Stilling well
4 Floating roof
5 Product
6 Tank