Operation – KROHNE OPTIWAVE 7300C EN User Manual
Page 60
6
OPERATION
60
OPTIWAVE 7300 C
www.krohne.com
03/2014 - 4000172405 - HB OPTIWAVE 7300 R07 en
...
Current changes
must be saved or
cancelled before
proceeding
Save, Cancel
3
Save
A.1.5
A.1.5
A.1.5
A.1.5
Outputs
Outputs
Outputs
Outputs
Follow this procedure to
Follow this procedure to
Follow this procedure to
Follow this procedure to
describe the output
describe the output
describe the output
describe the output
characteristics.
characteristics.
characteristics.
characteristics.
1
Output 1: Output
Function
Select an output function to
scale the current values
(output 1). This is not displayed
in the operator mode.
Level, Distance, Volume
(Mass), Ullage Volume
(Ullage Mass), Reflection
Level
2
Output 1 (HART):
This assigns a measurement
value to 4 mA (output 1).
min-max:
0…20 m / 0…65.62 ft
0 m / 0 ft
3
Output 1 (HART):
This assigns a measurement
value to 20 mA (output 1).
min-max:
0…90 m / 0…295.29 ft
Depends on
the output
function
4
Output Range
This sets the effective range of
output 1 with or without over-
run.
3.8…20.5 mA (NAMUR),
4…20 mA
4…20 mA
5
Error Handling
This sets the behaviour of
current output 1 if an error
occurs. Hold
Hold
Hold
Hold means that the
output current stays at the
value where the error
occurred. Hold
Hold
Hold
Hold is not available
if 3.8…20.5 mA (NAMUR) is the
output range.
3.6 mA, 22 mA, Hold
(4...20 mA range only)
22 mA
6
Output 1 (HART):
HART Address
Any HART
®
address greater
than 0 will activate HART
®
multidrop mode. Current
output stays constant at 4 mA.
min-max: 0…15
0
7
Output Function 2
4
Select an output function to
scale the current values
(output 2). This is not displayed
in the operator mode.
Level, Distance, Volume
(Mass), Ullage Volume
(Ullage Mass), Reflection
Level
8
Output 2 (Passive):
4
This assigns a measurement
value to 4 mA (output 2).
min-max:
0…20 m / 0…65.62 ft
0 m / 0 ft
9
Output 2 (Passive):
4
This assigns a measurement
value to 20 mA (output 2).
min-max:
0…90 m / 0…295.29 ft
Depends on
the output
function
10
OP2 Output Range
4
This sets the effective range of
output 2 with or without over-
run.
3.8…20.5 mA (NAMUR),
4…20 mA
4…20 mA
11
OP2 Error
Handling
4
This sets the behaviour of
current output 2 if an error
occurs. Hold means that the
output current stays at the
value where the error
occurred. Hold is not available
if 3.8…20.5 mA (NAMUR) is the
output range.
3.6 mA, 22 mA, Hold
(4...20 mA range only)
22 mA
12
Settings summary
Read only
Current changes
must be saved or
cancelled before
proceeding
Save, Cancel
Save
Menu
No.
Step
Function
Function description
Selection list
Default