Page 46: Keypad functions -2, Lcd messages -2, Key functions -2, Able 3-2 for, D 3-2 expla

background image

3-2

BHK-MG (OPR) 022014

3.2.1.4

KEYPAD FUNCTIONS

Keypad functions are listed in Table 3-2. Six keys have dual functions, depending on whether
the power supply is in command entry status (waiting for a command to be entered), or data
entry status (waiting for a number to be entered). Command entry status is indicated by a blink-
ing colon

(:_:_:)

and data entry status is indicated by a blinking equal sign

(=_=_=)

.

NOTE: Keys with dual functions are labeled with both a command and a number. The com-

mand label is referred to when the unit is in

(:_:_:)

command entry status; the num-

ber is referred to when the unit is in

(=_=_=)

data entry status.

TABLE 3-1. LCD MESSAGES

LOCATION

MESSAGE

DESCRIPTION

Top left

Loc/Rem

Digital control status: either Remote or Local

top right

CV/CC

Constant voltage mode/constant current mode

Bottom

left:

in command entry

n.nV

Actual output voltage

In data entry

(parameter)

e.g. OVset if OV SET key was pressed.

Bottom middle:

(:_:_:)

Command entry status

Top or Bottom middle:

(=_=_=)

Data entry status

Bottom

right:

in command entry

n.n mA

Actual output current.

In data entry

n.n V or n.n mA

Present programmed value of parameter, replaced by data entered.

TABLE 3-2. KEY FUNCTIONS

KEY

POWER SUPPLY

STATUS ACTIVE

DESCRIPTION

REFERENCE

PARAGRAPH

OUTPUT

ON/OFF

Command Entry

If top line of LCD reads Digital is OFF, press to enable the output. If output is on,
press to disable the output controlled by digital means (keypad or GPIB setting).
If analog control used, analog input must be turned off separately to disable the
output.

3.2.7

V SET

Command Entry

Press to set output voltage. After V SET is pressed, previous setting is dis-
played. Data entry required to enter new value of output voltage; press ENTER
to accept displayed value.

3.2.9

I SET

Command Entry

Press to set output current. After I SET is pressed, previous setting is displayed.
Data entry required to enter new value of output current; press ENTER to
accept displayed value.

3.2.9

LOCAL

Command Entry

If the power supply is in remote mode, keypad is disabled except for LOCAL
key. Press to enable keypad. If LCD reads KEYPAD LOCKED, the LOCAL key
is also disabled and can only be unlocked by remote operation.

3.2.4, B.84

RESET

Command Entry

Press to restore the power on default values: CV mode, output voltage = 0, out-
put current value chosen from the POWER-UP OR RESET CURRENT sub-
menu, overvoltage and overcurrent values per Table 1-2.

3.2.8