BHK-MG (OPR) (OPR) 022014

4-7

4.4.3

CALIBRATION PROCEDURE

1. Click on the Calibration Access button on the main panel (see Figure 4-2).This will open a

calibration panel (see Figure 4-3 and Table 4-1) that allows calibration in either voltage or
current mode. The Calibration Access button changes to Calibration Close.

2. The password has been set at the factory (see Table 4-2). Type the proper password in the

PASSWORD text box and press ENTER on the computer keyboard. If the password has
been forgotten, consult factory. If the password is correct, the VOLTAGE and CURRENT but-
tons will become active. Otherwise the message

ERROR! - Wrong Password

is displayed

and you must enter the correct password and press ENTER on the computer keyboard
again.

3. For Voltage Calibration, click VOLTAGE button. The text window reads

Remove Shunt

and

the message Connect DVM to Output is displayed above the SAVE CALIBRATION button.

4. Remove any connections between +OUT and –OUT. Connect DVM (high) to +OUT and

DVM (low) to –OUT, then click OK button.

NOTE: Use external DVM for all measurements.

5. The text window reads

Volt Min Calibration

and the message Measure and Adjust is

displayed beneath the text window. Using the " and # buttons (see PAR. 4.4.2), adjust the
output until the DVM is as close to 0 as possible but slightly positive, then click OK button.

TABLE 4-1. CALIBRATION PANEL FUNCTIONS

BUTTON OR WINDOW

Function

Text box (at top)

Used to enter unit password. Press ENTER on computer keyboard to vali-
date password.

VOLTAGE button

Initiates Voltage calibration

VOLTAGE Indicator below VOLTAGE button

Lights green when VOLTAGE calibration is acceptable. Lights red if VOLT-
AGE calibration needs to be done.

CURRENT button

Initiates Current calibration

CURRENT Indicator below CURRENT button

Lights green when CURRENT calibration is acceptable. Lights red if CUR-
RENT calibration needs to be done.

Message (as needed) below VOLTAGE and CUR-
RENT indicators

Provides instructions as needed about the calibration process.

White text window above OK button

Displays the parameter currently being calibrated.

!

button

Coarse Adjustment - Decreases the output value between five and 18 LSBs.

"

button

Coarse Adjustment - Increases the output value between five and 18 LSBs.

"

button

Fine Adjustment - Decreases the output value by one LSB.

#

button

Fine Adjustment - Increases the output value by one LSB.

OK button

Enters the value displayed in the white text window

SAVE CALIBRATION

Saves Calibration upon completion.

Message (as needed) above SAVE CALIBRATION
button

Provides instructions as needed about the calibration process.