beautypg.com

B.62 status:operation:enable command, Figure b-6. using status commands and queries, B.63 status:operation:enable? query – KEPCO BHK-MG 200W (Full Rack) Series User Manual

Page 145: B.64 status:operation[:event]? query, B.62, Stat, B.63, B.64, B.62, b.63, Stat:oper:enab

background image

BHK-MG (OPR) 022014

B-15

B.62 STATus:OPERation:ENABLe COMMAND

STAT:OPER:ENAB

Syntax:

Short Form:

STAT:OPER:ENAB 0 to 1313 (1 + 32 + 256 + 1024)

Long Form:

STATus:OPERation:ENABle 0 to 1313 (1 + 32 + 256 + 1024)

Description: Sets Operation Enable Register. The Operation Enable Register is a mask for enabling specific bits

in the Operation Event Register which will cause the operation summary bit (bit 7) of the Status Byte
register to be set Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inac-
tive, false). The operation summary bit is the logical OR of all the enabled bits in the Operation Event
register. (See example, Figure B-6.)

FIGURE B-6. USING STATUS COMMANDS AND QUERIES

B.63 STATus:OPERation:ENABLe? QUERY

STAT:OPER:ENAB?

Syntax:

Short Form: STAT:OPER:ENAB?

Long Form: STATus:OPERation:ENABle?

Return Value: 0 to 1313 (1 + 32 + 256 + 1024).

Description: Reads Operation Enable Register (see Table B-2). Returns value of Operation Enable Register

bits. Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inactive, false).
(See example, Figure B-6.)

B.64 STATus:OPERation[:EVENT]? QUERY

STAT:OPER?

Syntax:

Short Form: STAT:OPER[:EVEN]?

Long Form: STATus:OPERation[:EVENt]?

Return Value: 0 to 1313 (1 + 32 + 256 + 1024).

Description: Indicates changes in conditions monitored by Operational Event Register. Returns the value of

the Operation Event register. The Operation Event register is a read-only register which holds
(latches) all events that occur. Reading the Operation Event register clears it. (See example, Figure B-
6.)

NOTE:

The power supply is assumed to be operating in CV (constant voltage) mode.

OUTP ON

Output enabled.

VOLT 3;CURR 1E-2

Power supply programmed to 3V, 0.01A.

STAT:OPER:ENAB 1056

Mask enabled for CC, WTG and bits.

STAT:OPER:ENAB?

Returns 1056 (32 + 1024) (CC, WTG bits set).

STAT:QUES:ENAB 3

Mask enabled for OV and OC bits (1 + 2).

STAT:QUES:ENAB?

Returns 3 (1 + 2) indicating OV and OC bits are enabled.

STAT:PRES

Operation Condition Enable and Questionable Condition Enable registers
are preset.

INIT:CONT ON

Continuous triggers enabled.

STAT:OPER:COND?

Power supply returns 288 (256 + 32) to indicate that power
supply is constant voltage mode and Wait For Trigger is true.

--- MOMENTARILY SHORT POWER SUPPLY OUTPUT ---

STAT:OPER?

Returns 1312, e.g., indicating that since the last reading of the
Operation Event Register the power supply has entered
Constant Current mode, the Wait Trigger was set.

STAT:OPER?

Returns 0 indicating no changes since previous reading of the
Operation Event register.

STAT:QUES?

Returns 0 (no questionable conditions occurred since previous
reading

--- OVERTEMPERATURE CONDITION OCCURS

STAT:QUES?

Returns 8 (overtemperature condition tripped since the last
STAT:QUES? query).

STAT:QUES:COND?

Returns 8, (Power supply still in overtemperature condition).

STAT:QUES?

Returns 0, (Power supply no longer in overtemperature
condition).

STAT:QUES:COND?

Returns 8, (Power supply still in overtemperature condition).

SYST:ERR?

Power supply returns 0,“No error” message.