B.59 [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude]? QUERY

VOLT:TRIG?

Syntax:

Short Form: [SOUR:]VOLT[:LEV]:TRIG[:AMP]?
Long Form: [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude]?
Return Value: = digits with decimal point and Exponent, e.g., 2.71E+1 for 27.1

Description: Returns value representing voltage value to be programmed by *TRG command established by

VOLT:TRIG command). (See example, Figure B-1.)

B.60 [SOURce:]FUNCTion:MODE? QUERY

FUNC:MODE?

Syntax:

Short Form: FUNC:MODE?

Long Form: [SOURce:]FUNCtion:MODE?

Return Value: VOLT or CURR

Description: Identifies the operating mode of the power supply. VOLT = Constant Voltage mode (CV).

CURR = Constant Current mode (CC). (See example, Figure B-1.)

FIGURE B-5. PROGRAMMING VOLTAGE

B.61 STATus:OPERation:CONDition? QUERY

STAT:OPER:COND?

Syntax:

Short Form: STAT:OPER:COND?

Long Form: STATus:OPERation:CONDition?

Return Value: 0 to 1313 (1 + 32 + 256 + 1024).

Description: Returns the value of the Operation Condition Register (see Table B-2). The Operation Condition

Register contains unlatched real-time information about the operating conditions of the power supply.
Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inactive, false). (See
example, Figure B-6.)

NOTES: 1. The power supply is assumed to be operating in constant voltage (CV) mode.

2 Examples below are intended only to illustrate command functions. Refer to PAR. 3.3.6 for

programming techniques to optimize performance.

OUTP ON

Output enabled.

VOLT 218;CURR 1.1E-1

Power supply programmed to 218V, 0.11A.

VOLT 2.157E2

Power supply programmed to go to 215.7V.

VOLT?

Returns 2.157E2, indicating programmed voltage = 215.7V.

VOLT? MAX

For BHK 500-0.4MG, returns 500 (500V).

VOLT? MIN

Returns 0 (minimum allowable voltage).

VOLT:PROT 2.365E+2

Voltage protection set to 236.5V.

VOLT?

Returns 2.157E+2, indicating programmed voltage value = 215.7V.

VOLT:PROT?

Returns 2.365E+2, (programmed overvoltage protection = 236.5V

VOLT:PROT? MAX

For BHK 500-0.4MG, returns 5.5e+2 (550V).

VOLT 221;CURR 1.1E-1

Power supply programmed to go to 221V, 0.11A.

VOLT?

Returns 221.

VOLT:LIM[:HIGH] 300

Voltage limit set to 300V.

VOLT:LIM[:HIGH]?

Returns 300.

VOLT 333

Power supply not programmed to 333V., error message -301 posted.

VOLT?

Returns 2.21E2.

TABLE B-2. OPERATION CONDITION REGISTER, OPERATION ENABLE REGISTER,

AND OPERATION EVENT REGISTER BITS

CONDITION

NU

CC

NU

CV

NU

WTG

NU

CAL

BIT

15-11

10

9

8

7 - 6

5

4 - 1

0

VALUE

32,768 -

2048

1024

512

256

128 - 64

32

16 -2

1

CAL - INTERFACE COMPUTING NEW CALIBRATION

CONSTANTS

CC -

POWER SUPPLY IN CONSTANT CURRENT MODE

CV -

POWER SUPPLY IN CONSTANT VOLTAGE MODE

NU -

NOT USED

WTG WAIT FOR TRIGGER (SET BY INIT SUBSYSTEM)