beautypg.com

Jenway 3540 User Manual

Page 71

background image

8.4

Troubleshooting - Conductivity


Fault

Possible Cause

Action


No display

Check power supply

Check that correct 9V ac
power supply is connected and switched
on


Erratic display

Check power supply

The 3540 must be used with the
supplied 9V ac power supply
The 3540 will not operate with other
power supply units


Unstable display

Conductivity cell

Replace conductivity cell

defective

Display permanently Intermittent or no

Check cell connection to 3540

under or over range connection

Conductivity cell defective

Replace conductivity cell

Contaminated solutions

Replace solutions


Intermittent display

Conductivity cell not

Check connections

fitted

correctly


Reading drifts

Conductivity cell stored dry Allow to soak for 2 hours

CO

2

absorption by sample

Noticeable for low conductivity
- do not allow sample to stand in
unstoppered bottles


Non linear readings Conductivity cell not zeroed Zero conductivity cell

Poor reproducibility

Carryover between

Rinse cell in distilled water

solutions between

measurements


Will not print

IrDA connection broken

Line up units or use the supplied

RS232

connector

Paper out

The feed light on the printer will

flash

if

the

unit

requires

paper

Battery

flat

Connect

ac

power

supply


8.5

Functional Check – pH


The pH measurement function of the meter can be checked using the enclosed BNC shorting
cap (009 146).

1.

Remove the ATC probe, if connected.

2.

Set the manual temperature compensation to 25°C.

3.

Remove the pH electrode and replace it with the BNC shorting cap.

4.

Select the mV mode. The display should read ±1.


If the mV reading is greater then ±1mV perform a reset (refer Section x.x)

To make a measurement from this point, refit the ATC probe and pH electrode and calibrate
the 3540 using fresh buffer solutions.