3.30
Data Logging Set Up
The data logging set up menu enables all the functions of the data logger to be set,
including the event trigger, logging period and data destination. The function of the
Clear [CLR] key on stored data can also be set, there is also a sub-menu for the
Printer options.
Access the Data Logging Set Up menu by carrying out paragraphs 3.01 to 3.05, then
use the up or down arrow keys [ ] to move the highlight over the Data Logging Set
Up option, then press the Enter key [ ]. The Data Logging Set Up screen below will
be displayed…
Data Logging Set Up screen
3.31
Data Log Event
The data log event is the occurrence that triggers data to be sent to either internal
memory or the printer/computer via the serial port. There are eight options, making
this a useful and comprehensive function.
Disabled: Effectively switches data logging functions off.
Manual: Data logging is triggered from a press of the Store [STO] key only.
Timed Interval: Data logging is triggered on a period set by the Data Log interval.
Endpoint Detection: Data logging is triggered when the sample measurement
reaches stability – this can be used to enhance productivity when measuring batches
of samples by having the results automatically printed or stored without the operator
having to even press a button.
Timed after Endpoint: Data logging is triggered at the entered data log interval but
only after the endpoint detection – this enables data from reactions or slowly
changing sample streams to be logged with more control.
Alarm Set: Data logging is triggered when one of the alarm conditions is reached.
This is useful for tracking time/date of problems occurring.
Alarm Clear: Data logging is triggered when one of the alarm conditions is cleared.
This is useful for tracking time and date when problems are cleared.
Alarm Set & Clear: Data logging is triggered when one of the alarm conditions is set
and also when they are cleared. This gives useful data for analysing problems or