beautypg.com

Hermit 3000 channel setup, About external channels, Editing external channels – In-Situ Win-Situ 2000 Operators Manual User Manual

Page 25

background image

21

0024690 rev. 002 09/00

6
HERMIT 3000
CHANNEL SETUP

IN THIS SECTION . . .

• About External Channels
• Editing External Channels

ABOUT EXTERNAL CHANNELS

The HERMIT 3000 data logger can
accommodate up to eight 4-20 milliAmp
probes, such as pressure transducers,
attached to its eight numbered channel
connectors. To interpret the output of each
channel, Win-Situ requires specific information on the type of
probe connected and its calibration coefficients. If your software
supports the HERMIT 3000, you can supply this information by
“editing” each channel individually.

EDITING EXTERNAL CHANNELS

Use the following procedure to set up an external channel:

1. In the navigation tree for a HERMIT 3000, select a channel to which

you have attached (or plan to attach) a 4-20 mA probe.

2. In the Information pane, select the action button Edit Channel.

If the HERMIT is running a test, the Edit Channel function will
not be available.