GDS-806/810/820/840 Programming Manual

38

:GONogo:EXECute

(Available for firmware 2.0 and above only)

Start or stop the execution of GO/NO-GO comparison function. This command is

equal to the function key “F4” of GO/NO-GO menu.

Syntax

:GONogo:EXECute {0|1}

:GONogo:EXECute?

Arguments

0

→Stop GO/NO-GO comparison

1

→Start GO/NO-GO comparison

Returns

Note: Before running any Go/NoGo commands, the oscilloscope has to be initialized, please

add the command of “:GONogo:FUNCtion 1” in order to initial the oscilloscope.

:GONogo:FUNCtion

(Available for firmware 2.0 and above only)

This command is used to initialize or relieve the GO/NO-GO function. Before the

GO/NO-GO comparison is started. This command must be executed to enable the

comparison unit.

Syntax

:GONogo:FUNCtion {0|1}

:GONogo:FUNCtion?

Arguments

0

→Relieve GO/NO-GO function

1

→Initialize GO/NO-GO function

Returns