beautypg.com

GW Instek SFG-830 User Manual User Manual

Page 43

background image

p. 40 SFG-830

OPC Operation Complete Query

z Place an ASCII character 1 into the instrument‘s output queue when

all pending instrument operations are finished.

CLS

Clear Status Command

z This command clears status data structures. On the other hand, it

sets the bits of the〝Standard Event Register〞and the〝Status Byte
Register〞to zero.

z The Output Queue and the MAV bit will not be cleared.

ESE

Standard Event Status Enable Command

z *ESE, followed by a decimal value, sets the bits of the〝Standard

Event Status Enable Register〞

which correspond to that decimal value

to 1. This enables the assigned bits of the〝Standard Event Status
Register〞.

ESE Standard Event Status Enable Query

z This query asks for the contents of the 〝Standard Event Status Event

Register〞.The response is a decimal value, e.g.,〝255〞means all
events of the〝Standard Event Status Register〞are enabled, in other
word, all bits are 1.

ESR Standard Event Status Register Query

z Ask for the contents of the〝Standard Event Status Register〞. The

response is a decimal value. The query clears the register contents.