SFG-830 p.31

(5) Calibrating the Amplitude of Square Wave

Go through step

n

~

p

repeatedly to calibrate the output of Number 308 ~ 404.

n Set the Calibration Number.
o Use an oscilloscope with 50 ohm load to measure the output.
p Input an appropriate Calibration Value to make the output to be 10

±1Vp-p.

(6) Calibrating the Amplitude of Triangle Wave

n Set the Calibration Number to be 405.
o Use an oscilloscope at 50

Ω to measure the output.

p Input an appropriate Calibration Value to make the output to be 10

±1Vp-p.

(7) Calibrating the Symmetry of Square Wave

Go through step

n

~

p

repeatedly to calibrate the output of Number 406 ~ 502.

n Set the Calibration Number.
o Use a fixture with 50

Ω to measure the output.

s Input an appropriate Calibration Value to make the output to be as follows:

Calibration Number

Duty

406 ~ 437

50%

± 1% + 4ns

438 ~ 460

48%

461 ~ 486

45%

487 ~ 502

40%

(8) Finishing Manual Calibration

n Press [ STOR ] key.
ª “Default” will be shown on the display after 3 seconds.

(C) Automatic Calibration

n Press [Shift], [1], [8], [3], [0] keys to shift to automatic calibration mode.
ª In a few minutes “Auto Cal” will be shown on the display, indicates the

accomplishment of automatic calibration.

The table below lists the corresponding items to the Calibration Numbers:

Calibration Item

Calibration Number

Sine DC Gain

503

Square DC Gain

504

Triangle DC Gain

505

Adjusting DOB_U

506 ~ 602

All of the calibrating procedure is accomplished after the automatic calibration.