SFG-830 p.29

o Use [▲] and [▼] keys in MODIFY column or press numeric keys [0]~[9] to input

the desired value.

Example:
i.) If the display shows “00001 1000” and 00001 blinks, it means that the

Calibration Number is ready for input. Press FUNCTION [ ] key once, then
1000 becomes blinking and you can input the Calibration Value. Press
FUNCTION [ ] key once, then 00001 will blink again.

ii.) Press [4], [0], [6] and [Enter] keys will get to the Square Wave Symmetry

Calibration mode, and the Calibration Value is 2000. Press FUNCTION [ ] key
once, and 2000 will be blinking. Just input the desired figures, then the
symmetry of square wave will be changed.

The table below lists the corresponding Calibration Numbers to Calibration items:

Calibration Item

Calibration Number

Calibration Mode

Positive Attenuator Calibration

1 ~ 8

Negative Attenuator Calibration

9 ~ 16

Carrier Null Calibration

17 ~ 113

Amplitude of Sine Wave

114 ~ 307

Amplitude of Square Wave

308 ~ 404

Amplitude of Triangle Wave

405

Symmetry of Square Wave

406 ~ 502

Manual Calibration

Sine DC Gain Calibration

503

Square DC Gain Calibration

504

Triangle DC Gain Calibration

505

Offset Adjustment

506 ~ 602

Auto Calibration

The procedure of manual calibrating with software is as follows:

(1) Positive Attenuator Calibration

To calibrate the positive attenuator, you should go through Calibration Number 1 ~
8 in turn, and perform the following 3 steps for each number.
n Set the Calibration Number.
o Use a DMM (FLUKE-8842) at DCV range to measure the output.
p Key in an appropriate Calibration Value to make the output to be the particular

value listed below:

n

Calibration Number

p

Output should be :

1

39

± 0.1mV

2

78

± 0.5mV

3

156

± 1mV

4

312

± 1mV

5

625

± 2mV

6

1.25V

± 10mV

7

2.5V

± 10mV

8

o

Use a DMM (FLUKE-8842) at

DCV range to measure the

output

5V

± 10mV