7.9 Checking Frequency Accuracy

The deviation should be within

±5ppm.

n Set conditions:

Function : Sine Wave

Amplitude : 1Vp-p

Frequency : 10MHz

o Use a counter with 50 ohm load to measure the output.
p Check and make sure the output is 10MHz

± 40Hz.

7.10 Checking the Amplitude

(1) Check the Amplitude of Sine Wave

The deviation should be

±0.5dB (±5.9%).

n Set conditions:

Function : Sine Wave Amplitude : 3.54Vrms (10Vp-p)
Frequency : 100Hz

o Use a DMM (FLUKE-8842) at ACV range with 50 ohm load to measure the

output.

p Check and make sure that 3.33Vrms < output < 3.74Vrms.
q Keep the same Function and Amplitude, and set Frequency to be 1kHz, 10kHz,

100kHz, 1MHz, and 2 ~ 30MHz (step 2MHz) in turn.

r Perform step o~q (use measuring receiver HP-8902 or equivalent needed for

HF Amplitude check instead) in each frequency setting until the Frequency =
30MHz.

s Set conditions:

Function : Sine Wave

Amplitude : 1Vrms

Frequency : 100Hz

t Use a DMM (FLUKE-8842) at ACV range with 50 ohm load to measure the

output.

u Check and make sure that 0.94Vrms < output < 1.04Vrms.
v Keep the same Function and Amplitude, and set Frequency to be 1kHz, 10kHz,

100kHz, 1MHz, and 2 ~ 30MHz (step 2MHz) in turn.

w Perform step t~v (use measuring receiver HP-8902 or equivalent needed for

HF Amplitude check instead) in each frequency setting until the Frequency =
30MHz.

(2) Checking the Amplitude of Square Wave

The deviation should be

±12%.

n Set conditions:

Function : Square Wave

Amplitude : 10Vp-p

Frequency : 100Hz

o Use an oscilloscope with 50 ohm load to measure the output.
p Check and make sure that 8.8Vp-p < output < 11.2Vp-p.
q Keep the same Function and Amplitude, and set Frequency to be 1kHz, 10kHz,

100kHz, 1MHz, and 2 ~ 30MHz (step 2MHz) in turn.

r Perform step p~q in each frequency setting until the Frequency = 30MHz.