SFG-830 p.33

(3) Checking the Amplitude of Triangle Wave

The deviation should be

±5%.

n Set conditions:

Function : Triangle Wave

Amplitude : 10Vp-p

Frequency : 100Hz

o Use an oscilloscope with 50 ohm load to measure the output.
p Check and make sure that 9.5Vp-p < output < 10.5Vp-p.
q Keep the same Function and Amplitude, and set Frequency to be 1kHz, 10kHz,

and 100kHz in turn.

r Perform step p~q in each frequency setting until the Frequency = 100kHz.

(4) Check the Accuracy of DC Offset

The deviation should be

±1.5% + 1mV.

n Set conditions:

Function : Sine Wave

Amplitude : 0.0Vp-p

Frequency : 1kHz

DC Offset : 5V, -5V, 0V in turn

o Use a DMM at DCV range with 50 ohm load to measure the output.
p Check and make sure the corresponding output to each DC Offset is in the

range listed below:

DC Offset

Output

5V

4.925V

≤ output ≤ 5.075V

-5V

-5.075V

≤ output≤ -4.925V

0V

-1mV

≤ output ≤ +1mV.

(5) Checking the Sub Harmonic

The deviation should be <-50dBC.
n Set conditions:

Function : Sine Wave

Amplitude : 10Vp-p

DC Offset : 0V

o Use a spectrum analyzer to measure the output.
p Check and make sure that the corresponding output to each Frequency setting

is in the range listed below:

Frequency

Center Frequency of

Spectrum Analyzer

Harmonic

102kHz

51kHz

1.001MHz

501kHz

10.001MHz

5.001MHz

20.001mHz

10.001MHz

30.0mHz

15.0mHz

< -26.02dBm