p. 34 SFG-830

(6) Checking the Harmonic Distortion

The deviation should be :

Deviation

Frequency

< -50dBC

< 100kHz

< -40dBC

0.1 ~ 1MHz

< -30dBC

1 ~ 10MHz

< -25dBC

10 ~ 30MHz

n Set conditions:

Function : Sine Wave

Amplitude : 1Vp-p

Frequency : 100Hz

o Use a distortion meter (DM-155A/B) with 50 ohm load to measure the output.
p Check and make sure that the output should be < -50dBC (0.31%).
q Set the Frequency to be 1kHz, 10kHz, 50kHz, and 100kHz in turn.
r Perform step p and q until the Frequency = 100kHz.
s Use a spectrum analyzer to measure the output.
t Check and make sure that the corresponding output to each Frequency setting

is in the range listed below:

Frequency

Output

500kHz

< -40dBC

5MHz

< -30dBC

15MHz

< -25dBC

30MHz

< -25dBC

(7) Checking the Phase Noise

The deviation should be <-50dBC.
n Set conditions:

Function : Sine Wave

Amplitude : 13dBm

Frequency : 10MHz

o Use a spectrum analyzer to measure the output.
p Check and make sure that the difference between the powers of 10MHz and

10MHz+15kHz is less than 50dBC.

(8) Checking the Rise Time of Squire Wave

The deviation should be : Rise Time < 15ns and Overshoot < 5% at full scale
output.
n Set conditions:

Function : Square Wave

Amplitude : 10Vp-p

Frequency : 1MHz

o Use an oscilloscope with 50 ohm load to measure the output.
p Check and make sure that the Rise Time < 15ns and

the Overshoot / Undershoot < 500mVp-p.