GDS-2000 Series Programming Manual

73

:TEMPlate:POSition:MIN

Sets and query the position of minimum template. This command will not change

the position value that stored in internal memory, unless you save the template by

the command of :TEMPlate:SAVe:MINimum. That means that if the
oscilloscope is re-opened again without present template storage, the oscilloscope

will recall the previous stored position value from internal memory automatically.

Before running this command, the oscilloscope has to be initialized as “Normal

Template” mode, please add the command of “:TEMPlate:MODe 0” in order to
change the settings.

Syntax

:TEMPlate:POSition:MIN

:TEMPlate:POSition:MIN?

Arguments

is the desired template position. The range is from -12.0 ~ +12.0 div.

Returns

Div

Note: 1 division = 25 dots

Note: Before input any template editing commands, the oscilloscope has to be initialized,

please add the command of “:GONogo:FUNCtion 1” in order to initial the oscilloscope.