beautypg.com

GW Instek GDS-2000 series Programming manual User Manual

Page 49

background image

GDS-2000 Series Programming Manual

49

:HARDcopy:MODe

Select the hardcopy output style. The entire LCD screen, waveform raw data,

oscilloscope settings can be saved on flash disk or print out the entire LCD screen

to mostly USB printers.

Syntax

:HARDcopy:MODe {0|1|2}

:HARDcopy:MODe?

Arguments

0→ Save image

1→ Save all

2→ USB Printer

Returns

:HARDcopy:RATio

Select the print-out image size for hardcopy function. The adjustable range is from

10% to 100%. The one hundred percent will fill the entire image with selected

printer’s paper size; ten percent will reduce 90% of image size with original selected

printer’s paper size. The default setting is 50%.

Syntax

:HARDcopy:MODe

:HARDcopy:MODe?

Arguments

between 10~100

Returns

%