PSM-SERIES PROGRAMMABLE POWER SUPPLY

USER MANUAL

⎯ ⎯

49

6-4. Adjustment and Calibration

--Preparation

1. 30 minutes warm up before calibration.

2. Ambient temperature:23±5°C, Humidity: Under RH80%.

--Output Calibration Steps:

[Step 1]

Press [SHIFT][ .] to appear Password input window, input Password (vary

with different models: PSM-2010= 2010, PSM-3004 = 3004, PSM-6003 =

6003) by using the number key, press [ENTER] key

[Step 2]

When get into Calibrated item selection window, set the cursor to the

calibrated item by using the knob and press [ENTER]. After the calibrated

item is selected, proceeding voltage, current and over-voltage calibration

step by step.

[Step 3] Voltage Calibration Steps

Set to [VOLTAGE] by using the knob and press [ENTER] getting into

voltage calibration procedure.

PSM-SERIES PROGRAMMABLE POWER SUPPLY

USER MANUAL

⎯ ⎯

50

[Step 3.1]

Firstly, proceed the calibration procedure of LO calibrated point, input

DMM measurement value, then press [ENTER].

Note: The DMM selected for the measurement must have the resolution of

4 digits after the decimal point at least, and take three digits after the

decimal point to be the effective value and run off the rest.

[Step 3.2]

Next, proceed the calibration procedure of MI calibrated point, input DMM

measurement value, then press [ENTER].

Note: The DMM selected for the measurement must have the resolution of

4 digits after the decimal point at least, and take three digits after the

decimal point to be the effective value and run off the rest.

[Step 3.3]

Finally, proceed the calibration procedure of HI calibrated point, input

DMM measurement value, then press [ENTER].