beautypg.com

GW Instek PSM-Series User Manual User Manual

Page 26

background image

PSM-SERIES PROGRAMMABLE POWER SUPPLY

USER MANUAL

⎯ ⎯

45

6. MAINTENANCE

The following instructions are executed by qualified personnel only. To

avoid electrical shock, do no perform any servicing other than the operating

instructions unless you are qualified to do so.

6-1. Fuse Replacement

If the fuse blows, the display will not light and the power supply will not

operate. The fuse should not normally open unless a problem has

developed in the unit. Try to determine and correct the cause of the blown

fuse, then replace only with a fuse of the correct rating and type.

WARNING. For continued fire protection. Replace fuse

only with 250V fuse of the specific type and rating, and

disconnect power cord before replacing fuse.

6-2. Line Voltage Conversion

The primary winding of the power transformer is tapped to permit

operation from 100, 120, 220, or 230VAC, 50/60 Hz line voltage.

Conversion from one line voltage to another is done by change AC

selects switch.

The rear panel identifies the line voltage to which the unit was factory

set. To convert to a different line voltage, perform the following

procedure:

WARNING

PSM-SERIES PROGRAMMABLE POWER SUPPLY

USER MANUAL

⎯ ⎯

46

(1) Make sure the power cord is unplugged.

(2) Change the AC selects switch to the desired line voltage position.

A change in line voltage may also require a corresponding change of

fuse value. Install the correct fuse value as listed on rear panel.

6-3. Measurement Techniques

For an accurate and convenient measurement, please set up the instrument

according to the following:

1) Setup:

Figure 6-1