PSM-SERIES PROGRAMMABLE POWER SUPPLY

USER MANUAL

⎯ ⎯

47

2) The calibrated devices:

For general calibration, user can select the suitable calibration devices

by their own. Please refer to the below recommendation:

A. Electronic Load

In order to meet the demand of the auto calibration, the Electronic

Load with GPIB is well recommended. The main function of the

Electronic Load is to control the short test (CC current calibration),

the connection test (voltage load regulation, transient response

time) and open test( CV voltage calibration).

Measurement Techniques

Assure a precise Load regulation, Voltage peak to peak, and

transient response time, the test joint point must be configured to

sampling terminals of S+ and S-.

B. Current Monitoring Resistor

The resistor must have sufficient Watt value that can meet the

calibration need to control the change amount less than 15mA

when a 20.00A current run over it for a long period.

C. DVM or RMS Voltmeter

The meter used for the output voltage measurement should have

0.1mV resolution (accuracy: 0.01%). For the output current

measurement, the meter should have 0.001mV resolution(accuracy:

0.01%) to measure the voltage of the shunt, then transfer it to the

corresponding current value.

PSM-SERIES PROGRAMMABLE POWER SUPPLY

USER MANUAL

⎯ ⎯

48

D. Oscilloscope

To test the voltage noise and output oscillator when the power supply

is in the CV and CC mode.

E. Person Computer (PC)

This is a supplementary tool for the auto calibration including interface

card and auto calibration software of its own.