PSM-SERIES PROGRAMMABLE POWER SUPPLY

USER MANUAL

⎯ ⎯

19

When the current load runs through output terminal, if the current

exceeds the setting value, the operation of the instrument will be set to

Constant Current Mode (C. C. Mode), if the current does not exceed the

setting value, the operation will be set to Constant Voltage mode (C. V.

Mode).

z OVP SET: Over Voltage protection setting (LEVEL, ON/OFF

/CLEAR)

Setting method: After pressing [OVP SET], get into Over Voltage

Protection Level setting, now use the number key (or knob) to input

setting value, then press [ENTER] to complete the setting change.

When get into Over Voltage protection state setting by selecting Over

Voltage protection state and pressing [ENTER], use the knob to select

ON/OFF/CLEAR, then press [ENTER] to complete the setting change.

Remark: When the Over Voltage protection of the instrument is

triggered, use the function of Over Voltage Protection

Clear (OVP CLEAR) to clear its protection state.

For example: Set Over Voltage protection value to 21.00V at ON state.

After pressing [OVP SET] to get into OVP LEVEL setting, then input

[2][1][.][0][0][ENTER] to get into OVP STATE setting, now set the

knob select state to ON and press [ENTER] to complete setting.

PSM-SERIES PROGRAMMABLE POWER SUPPLY

USER MANUAL

⎯ ⎯

20

z OCP SET: Over current protection setting (LEVEL, ON/OFF/

CLEAR, DELAY)

Setting method: After pressing [OCP SET] to get into Over Current

Protection Level (OCP LEVEL) setting, use the number key (or knob) to

input setting value and press [ENTER] to complete the setting change.

When get into over current protection state, use the knob to select

ON/OFF/CLEAR, then press [ENTER] to change the setting.

When the Over Current Protection is set to ON, then get into Over

Current Protection Delay setting (The minimum Delay time is at 0

second, the maximum is at 10second, and the scale base is 0.1 second).