PSM-SERIES PROGRAMMABLE POWER SUPPLY

USER MANUAL

⎯ ⎯

35

4) Set the desired output current value:

Press [I SET], the panel operation is set to current value input

mode, change to the desired value by using the number key or the

wheel knobs (the wheel knob can be used together with [<] or [>]

to adjust the resolution.

5) Start the output:

Press [OUTPUT] to enable output, now, the meter displays the

actual output measurement value.

6) Constant current mode confirmation

Check whether the C.C. indicator is on or not to make sure the

output operation is under the constant current mode. If the C.V.

indicator is on, it needs to enlarge its voltage limit value to assure

that the output operation is under constant current mode.

z

Programming Over Voltage Protection, OVP

Set OVP level and the protection circuit of the OVP status.

1) Input the value of the Over Voltage protection limit value and set

the OVP status.

2) Confirm the function of OVP (Over Voltage Protection):

Adjust the output voltage up near to OVP level to confirm the

OVP function, when the OVP circuit is triggered, the OVP loop

will drop the output voltage near to zero, now, the OVP indicator

will be blinking.

3) Clear the OVP function:

PSM-SERIES PROGRAMMABLE POWER SUPPLY

USER MANUAL

⎯ ⎯

36

Firstly, remove the external test unit, such as battery, from the

instrument, adjust the voltage down or raise up the OVP level,

then clear the OVP.

z

Programming Over Current Protection, OCP

Set OCP level and the protection circuit of the OCP status.

1) Input the value of the Over Voltage protection limit value and set

the OCP status.

2) Confirm the function of OCP (Over Current Protection):

Adjust the output current up near to OCP level to confirm the

OCP function, when the OCP circuit is triggered, the OCP loop

will drop the output current near to zero, now, the OCP indicator

will be blinking.

3) Clear the OCP function:

Firstly, remove the external test unit, such as battery, from the

instrument, adjust the output current down or raise up the OCP

level, then clear the OCP.

z

Storing and Recalling Operating:

1) The storing setting function including the store of the Output

range, Output voltage value, Output current value, Over Voltage

protection level, Over Current protection level, Over Voltage

protection status, Over Current protection status and the Delay

time.