PSM-SERIES PROGRAMMABLE POWER SUPPLY
USER MANUAL
⎯ ⎯
21
z OCP DELAY TIME CHART
Remark: When the over current protection of the instrument is
triggered, use the function ++of Over Current Protection
Clear (OCP CLEAR) to clear its protection state.
PSM-SERIES PROGRAMMABLE POWER SUPPLY
USER MANUAL
⎯ ⎯
22
For example: Set over current protection value to 21.00A at ON state.
After pressing [OCP SET], get into OCP LEVEL setting, input
[2][1][.][0][0][ENTER], now set the knob select state to ON and press
[ENTER] to complete setting.
z Output voltage/current selection
When re-power on, the output range is stayed at the previous setting
before power off. Besides, whatever the output is at ON or OFF status,
the output range can be switched over dynamically. When the setting
value is more than the value of the switched voltage/current range, it
will be modified to maximum value of the switched range automatically.
If the setting value is less than the value of the switched voltage/current
range, it will be stayed as it is.
z PSM-2010 :20V,10A/8V,20A
z PSM-3004 :30V,4A/15V,7A
z PSM-6003 :60V,3.3A/30V,6A
For example: Set output range of PSM-2010 to 20V, 10A
Press [SHIFT][20V,10A], the output range will be changed to 20V,10A.