GW Instek PSM-Series User Manual User Manual
Page 20
PSM-SERIES PROGRAMMABLE POWER SUPPLY
USER MANUAL
⎯ ⎯
33
5-3. Operation Method
The applied Voltage/Current Unit for this series instruments is Volt and
Amp.
z
Constant Voltage Operation
1) Connect load to output terminal:
For the safety, when connect the load to output terminals of (+)
and (-), it must turn off the power.
2) Select output range:
Turn on the power after the load is well connected, select the
adequate operation range by pressing [SHIFT][8V/20A] or
[SHIFT][20V/10A]).
3) Set the current limit value:
Press [I SET], the panel operation is set to current value input
mode, set the desired change value by using the number key or
the wheel knobs (the wheel knob can be used together with [<] or
[>] to adjust the resolution).
4) Set the desired output voltage value:
Press [V SET], the panel operation is set to voltage value input
mode, change to the desired value by using the number key or the
wheel knobs (the wheel knob can be used together with [<] or [>]
to adjust the resolution).
PSM-SERIES PROGRAMMABLE POWER SUPPLY
USER MANUAL
⎯ ⎯
34
5) Start the output:
Press [OUTPUT] to enable output, now, the meter displays the
actual output measurement value.
6) Constant voltage mode confirmation
Check whether the C.V. indicator is on or not to make sure the
output operation is under the constant voltage mode. If the C.C.
indicator is on, it needs to enlarge its current limit value to assure
that the output operation is under constant voltage mode.
z
Constant Current Operation
1) Connect load to output terminal:
For the safety, when connect the load to output terminals of (+)
and (-), it must turn off the power.
2) Select output range:
Turn on the power after the load is well connected, select the
adequate operation range by pressing [SHIFT][8V/20A] or
[SHIFT][20V/10A]).
3) Set the voltage limit value:
Press [V SET], the panel operation is set to voltage value input
mode, set the desired change value by using the number key or
the wheel knobs (the wheel knob can be used together with [<] or
[>] to adjust the resolution.