[SOURce:]VOLTage[:LEVel]:TRIGgered
[:AMPLitude]
Set
Query
Description
Sets or queries the voltage level in volts when a
software trigger has been generated.
Syntax
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]
{
Query Syntax
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]?
[MIN|MAX]
Parameter/Return
0%~105% of the rated voltage output in volts.
MIN
Minimum current level.
MAX
Maximum current level.
Example
SOUR:VOLT:LEV:TRIG:AMPL 10
Sets the voltage level to 10 volts when a software
trigger is generated.
[SOURce:]VOLTage:PROTection[:LEVel]
Set
Query
Description
Sets or queries the overvoltage protection level.
Syntax
[SOURce:]VOLTage:PROTection[:LEVel]
{
Query Syntax
[SOURce:]VOLTage:PROTection[:LEVel]? [MIN|MAX]
Parameter/Return
OVP range in volts.
MIN
Minimum OVP level
MAX
Maximum OVP level
Example
SOUR:VOLT:PROT:LEV MAX
Sets the OVP level to its maximum.
[SOURce:]VOLTage:SLEW:RISing
Set
Query
Description
Sets or queries the rising voltage slew rate. This is
only applicable for CV slew rate priority mode.