beautypg.com

Initiate commands, Initiate[:immediate]:name – GW Instek PSW-Series Programming & Command manual User Manual

Page 58

background image

PSW Series Programming Manual

58

Parameter

0

Turns blink OFF

OFF

Turns blink OFF

1

Turns blink ON

ON

Turns blink ON

Return parameter 0

Turns blink OFF

1

Turns blink ON

Example

DISP:BLIN 1
Turns blink ON.

Initiate Commands

INITiate[:IMMediate]:NAME ............................................. 58

INITiate[:IMMediate]:NAME

Set

Description

The INITiate command starts the TRANsient or
OUTPut trigger.
See the trigger commands on page 75 for usage
details.

Syntax

INITiate[:IMMediate]:NAME {TRANsient|OUTPut}

Parameter

TRANSient

Starts the TRANsient trigger.

OUTPut

Starts the OUTPut trigger.

Example

INITiate:NAME TRANient
Starts the TRANSient trigger.