beautypg.com

GW Instek PSW-Series Programming & Command manual User Manual

Page 106

background image

PSW Series Programming Manual

106

Error Code

Description

-100 Command
Error

This is the generic syntax error for devices that
cannot detect more specific errors. This code
indicates only that a Command Error as
defined in IEEE 488.2,11.5.1.1.4 has occurred.

-102 Syntax error

An unrecognized command or data type was
encountered; for example, a string was received
when the device does not accept strings.

-103 Invalid
separator

The parser was expecting a separator and
encountered an illegal character; for example,
the semicolon was omitted after a program
message unit,
MEAS:VOLT:DC?:MEASCURR:DC?

-104 Data type
error

The parser recognized a data element different
than one allowed; for example, numeric or
string data was expected but block data was
encountered.

-108 Parameter
not allowed

More parameters were received than expected
for the header; for example, the KLOCk
command only accepts one parameter, so
receiving SYSTem:KLOCk 1,0 is not allowed.

-109 Missing
parameter

Fewer parameters were recieved than required
for the header; for example, the KLOCk
command requires one parameter, so receiving
KLOCk is not allowed.

-111 Header
separator error

A character which is not a legal header
separator was encountered while parsing the
header; for example, no white space followed
the header, thus APPL5,1 is an error.