beautypg.com

Remote control – GW Instek PSW-Series Programming & Command manual User Manual

Page 103

background image

REMOTE CONTROL

103

Status Byte Register & Service Request Enable Register

Overview

The Status Byte register consolidates the status
events of all the status registers. The Status Byte
register can be read with the *STB? query and
can be cleared with the *CLS command.

ERR

QUES

MAV

ESB

MSS

OPER

0

1

2

3

4

5

6

7

Output

Buffer

:

:

:

Error Que

:

:

:

Status

Byte

Register

0

1

2

3

4

5

6

7

&

&

&

&

&

&

&

*SRE

*SRE?

Service

Request

Enable

Register

Logical OR

*STB?

From Operation

Status Register

From

Questionable

Status Register

From Standard

Event Status

Register

Bit Summary

Event

Bit #

Bit
Weight

ERR (Error Event/Queue)

If data is present in the Error
queue, the ERR bit will be set.

2

4

QUES (Questionable Status

Register)
The summary bit for the
Questionable Status Register
group.

3

8

MAV (Message Available) This is

set when there is data in the
Output Queue waiting to be read.

4

16