beautypg.com

Standard event status register group, Remote control – GW Instek PSW-Series Programming & Command manual User Manual

Page 101

background image

REMOTE CONTROL

101

Standard Event Status Register Group

Overview

The Standard Event Status Register Group
indicates if any errors have occurred. The bits
of the Event register are set by the error event
queue.

OPC

RQC

QUE

DDE

EXE

CME

URQ

PON

Event

Enable

0

1

2

3

4

5

6

7

0

1

2

3

4

5

6

7

Standard Event Status Register

&

&

&

&

&

&

&

&

To Status Byte Register

*ESR?

Logical OR

*ESE

*ESE?

Bit Summary

Event

Bit #

Bit
Weight

OPC (Operation complete)

The OCP bit is set when all
selected pending operations are
complete. This bit is set in
response to the *OPC command.

0

1

RQC (Request control)

1

2

QUE (Query Error)

The Query Error bit is set in
response to an error reading the
Output Queue. This can be
caused by trying to read the
Output Queue when there is no
data present.

2

4

DDE (Device Dependent Error)

Device specific error.

3

8